TGA Action ID RC-2026-RN-00102-1
Type of Product Medical Device
Product Name/Description Ventilator Breathing Circuit for Non-invasive Ventilation Devices

Catalog Number: AMVC1792-156

Lot Numbers: 021025, 160924, 230725, 181024
ARTG Number 475290
Action Type Recall / Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action The manufacturer, Armstrong Medical, has received reports from customers describing holes and tears between the tubing and the purple cuff connector (of ventilator circuits and anaesthetic circuits). These defects have resulted in delays in treatment and required replacement of the breathing circuit. Following internal investigation, a manufacturing defect has been identified that may cause holes or tearing at the junction between the tubing and the purple cuff connector. This can occur during handling, including when replacing a filter.

No patient harm has been reported due to this problem.
Instructions for Customers/Users Customers are to:

1. Identify and quarantine all affected lots listed in Table 1 of the Customer Letter.
2. Cease use immediately.
3. Return all affected units following the instructions provided by your Breathing and Medical representative.

However, where a suitable alternative product cannot be sourced, continued use of the product is permitted in adherence with the following controls outlined in the Customer Letter.

This action was finalised on 28/10/2026.
Action Commencement Date 18/02/2026
Responsible Entity Breathing and Medical
Contact Information 1800 335 333 - Breathing and Medical