TGA Action ID RC-2026-RN-00093-1
Type of Product Medical Device
Product Name/Description MediHoney Wound Products

Product Descriptions: MediHoney Medical Honey 50g Tube - Sterile, MediHoney Barrier Cream 50g Tube, MediHoney Alginate Dressing 2" x 2" (box of 10), MediHoney Apinate Dressing 10 x 10cm 5 Pack, MediHoney Tulle Dressing 10 x 10cm 5 pack

Catalogue Numbers: 405, 582, 794, 795, 796

All unexpired lot numbers

ARTG 178475 & 178728 (Cancelled as of 11/08/2025)
ARTG Number
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Packaging failures were identified related to the MediHoney Wound products, which could lead to a breach in the sterile barrier. Two potential problems were identified:

1. Inadequate sealing of the foil membrane
2. Packaging seal is folded resulting in a channel for product to leak.

There have been no reported complaints in Australia or New Zealand for the MediHoney Wound products.
Instructions for Customers/Users Customers are to:
- Identify the products, cease use and quarantine them.
- Distributors are to contact their customers to request return of affected stock.
- Distributors are to then consolidate the stock returned from their customers, for return to Integra.

A credit note will be processed upon receipt and verification of returned goods by 30 April 2026.

This action has been closed out on 09/09/2026
Action Commencement Date 11/02/2026
Responsible Entity Integra Neurosciences Pty Ltd
Contact Information 1300 550 599 - Integra Neurosciences Customer Service