| TGA Action ID |
RC-2026-RN-00091-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips IntelliVue Patient Monitors
Product Descriptions: M8102A M8105A M8001A M8002A M8004A M8007A M3002A 8
Product Codes: M8102A, M8105A, M8001A, M8002A, M8004A, M8007A, M3002A, 867030, 866060, 866062, 866064, 866066, 865241, 866471, 865240, 866470 |
| ARTG Number |
94238, 213953, 400694, 423852 |
| Unique Device Identifier (UDI) |
00884838000278, 00884838000230, 00884838000193, 00884838000155, 00884838000179, 00884838000209, 00884838000261, 00884838082588, 00884838038752, 00884838038769, 00884838038776, 00884838038783, 00884838029088, 00884838083332, 00884838020733, 00884838070790 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class I |
| Reason for the Action |
Philips has received complaints that the IntelliVue monitors did not alarm. In some cases, these complaints alleged delayed patient treatment due to the monitors not alarming; and, in some cases, there were reports of severe patient harm or death. It was determined that in these instances the Alarms had been switched off for an infinite period, and the user may have been unaware that the alarms were off. |
| Instructions for Customers/Users |
Philips is informing customers of the observed errors and related hazards of enabling infinite Alarms Off, including providing instructions on how to configure the device alarm settings. Philips recommends reviewing the configuration of the monitors with infinite Alarms Off functionality enabled and reconfigure settings to require additional confirmation to ensure users are aware of their action when activating infinite Alarms Off.
For IntelliVue MP5 models only: Philips has developed a software update which will enforce the additional confirmation step when the user enables infinite Alarms Off. To update the MP5 software to include this feature, please reach out to your local Philips representative.
Philips is also providing an educational video on this topic. |
| Action Commencement Date |
4/02/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |