| TGA Action ID |
RC-2026-RN-00085-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
VITROS XT 3400/3600/4600/5600/XT 7600 Systems. An in vitro diagnostic medical device (IVD).
Catalog Numbers: 684 4458, 680 2783, 680 2914, 680 2445, 690 0440, 680 2413, 680 2915, 684 4461, 627 2222
All Lot Numbers |
| ARTG Number |
180199 |
| Unique Device Identifier (UDI) |
10758750031986, 10758750002979, 10758750007103, 10758750012343, 10758750033201, 10758750002740, 10758750007110, 10758750031610, 10758750035656 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
QuidelOrtho has identified a problem where a combination of a software anomaly and use error can lead to an erroneous result generated using an unintended reagent cartridge for a different assay.
The erroneous test result may or may not be clinically plausible. An erroneous patient test result may lead to an inappropriate diagnosis and physician intervention.
QuidelOrtho has received a complaint regarding this problem. No adverse events have been reported. |
| Instructions for Customers/Users |
When performing a Manual Load procedure, customers are to confirm that the intended reagent to be loaded matches what is entered on the Manual Load Cart or Pack dialogue box. During a Manual Load procedure, if a VITROS MicroSlide, MicroTip, or MicroWell reagent is presented after opening the LOAD DOOR, that reagent must be unloaded, and the correct, intended reagent must be loaded into that slot.
QuidelOrtho will inform customers when a software update is available to resolve this problem. |
| Action Commencement Date |
2/02/2026 |
| Responsible Entity |
|
| Contact Information |
1800 032 359 - QuidelOrtho Technical Services Centre |