TGA Action ID RC-2026-RN-00076-1
Type of Product Medical Device
Product Name/Description MRDSET-NSW (Mechanical Restraint Device Set)

Code 41720

Batch Number: B03
ARTG Number 446531
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action Due to a buckle failure, it was found that a Mechanical Restraint Device Set was able to be detached by an agitated patient, and a staff member at the hospital was injured.

Given the risk to patient and staff safety, this action was undertaken prior to TGA notification (Immediate Market Action as defined within the PRAC).
Instructions for Customers/Users Customers were contacted in December 2025 and notified to isolate all stock of batch B03.

Customers should continue to identify and quarantine affected stock of Batch B03, and subsequently return to the sponsor.

To ensure the safety of ambulance staff and patients, the sponsor has supplied kits with new hardware from a new supplier.

This action has been closed out on 17/08/2026
Action Commencement Date 4/02/2026
Responsible Entity Hemco Industries Pty Ltd