| TGA Action ID |
RC-2026-RN-00076-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
MRDSET-NSW (Mechanical Restraint Device Set)
Code 41720
Batch Number: B03 |
| ARTG Number |
446531 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class I |
| Reason for the Action |
Due to a buckle failure, it was found that a Mechanical Restraint Device Set was able to be detached by an agitated patient, and a staff member at the hospital was injured.
Given the risk to patient and staff safety, this action was undertaken prior to TGA notification (Immediate Market Action as defined within the PRAC). |
| Instructions for Customers/Users |
Customers were contacted in December 2025 and notified to isolate all stock of batch B03.
Customers should continue to identify and quarantine affected stock of Batch B03, and subsequently return to the sponsor.
To ensure the safety of ambulance staff and patients, the sponsor has supplied kits with new hardware from a new supplier.
This action has been closed out on 17/08/2026 |
| Action Commencement Date |
4/02/2026 |
| Responsible Entity |
|
| Contact Information |
1300 065 087 - Patricia Warke, Hemco Health Equipment |