TGA Action ID RC-2026-RN-00072-1
Type of Product Medical Device
Product Name/Description STA Liatest Free Protein S 6. An in vitro diagnostic medical device (IVD)

Reference Number: 00516

Lots: 271 971, 272 304, 272 529, 272 859, 273 018, 273 426, 273 635
ARTG Number 185169
Unique Device Identifier (UDI) (01)03607450005165(11)241001(17)260131(10)271971(241)00516, (01)03607450005165(11)241114(17)260228(10)272304(241)00516, (01)03607450005165(11)241218(17)260331(10)272529(241)00516, (01)03607450005165(11)250221(17)260531(10)272859(241)00516, (01)03607450005165(11)250306(17)260630(10)273018(241)00516
Action Type Recall / Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action Following customer complaints, Stago confirmed out-of-range results for the STA Liatest Control N and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

The root cause investigations revealed a problem linked to a single lot of one of the biological raw material components where effects on results become apparent gradually over time.

The potential risk associated with this underestimation would be an error in classifying protein S deficiency: a patient with type II deficiency (qualitative defects) could be incorrectly classified as type I or III (quantitative defects). However, the risk remains negligible, given the rarity of type II deficiencies and the fact that this classification has no impact on patient care.
Instructions for Customers/Users Customers are to:
- Discontinue use and dispose of lots with a short expiration date (Lots 271971, 272304, 272529, 272859, 273018).

- Stago will issue new calibration data for lots with a later expiration date (Lots 273426, 273635).
Action Commencement Date 3/02/2026
Responsible Entity Diagnostica Stago Pty Ltd
Contact Information 03 9840 555 - Stago