Print version of this report
TGA Action ID RC-2026-RN-00071-1
Type of Product Medical Device
Product Name/Description Philips Azurion R4.0 Systems

Affected Systems: Azurion 7 M12, Azurion 7 M20, Azurion 7 M12, Azurion 7 M20, Azurion 5 M12, Azurion 5 M20

Model Numbers: 722078, 722079, 722223, 722224, 722227, 722228
ARTG Number 225815, 370112
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified five software problems affecting Azurion R4.0 systems that may result in loss of imaging (X-ray) functionality, and/or loss of motorised movement, and/or incorrect image content, and/or loss of data.

Problem 1 – System keeps restarting
Problem 2 – AMC triple drive - Due to a firmware problem movements of the stand might be slow or might become unavailable
Problem 3 - System remains in continuous restart mode after the start-up
Problem 4 – Misalignment of Marker Tool Overlay – Applicable only to Azurion systems with the Marker Tool option
Problem 5 - Longitudinal Position Error – Applicable only to Azurion systems with Poly-G3 frontal stand - the system may detect a mismatch between the expected longitudinal (set-) position and the measured (actual) position of the stand. When the mismatch is detected, the motion power to the frontal stand is automatically cut off.
Instructions for Customers/Users Customers should read the customer letter that outlines all the steps to mitigate these problems whilst awaiting a fix, which may include a cold restart.
Philips expects to have software update 4.0.5, that resolves problems 1 - 3, released by Q2 2026 (subject to regulatory clearance).
The Misalignment of Marker Tool Overlay problem 4) and Longitudinal Position Error (problem 5) will not be resolved with software update 4.0.5. The solution for those problems is planned for Q1 2027 (subject to regulatory clearance) with software update 4.0.10.

UPDATE 11/8/2026: Philips had previously communicated in Feb 2026 that problems 1-3 would be corrected through software update R4.0.5, expected to be available in Q2 2026, and that problems 4-5 would be corrected through software update R4.0.10, expected to be available in Q1 2027.
Philips notes that software update R4.0.5 was available in July 2026 and now expects software update R4.0.10 by March 2027 (subject to applicable regulatory clearances).
Action Commencement Date 3/02/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team