TGA Action ID RC-2026-RN-00060-1
Type of Product Medical Device
Product Name/Description Prismaflex Sets

Prismaflex ST100 Set
Product Code: 107636
All lots including and manufactured after 24F0069

Prismaflex ST150 Set
Product Code:107640
All lots including and manufactured after 24E0059

Prismaflex TPE2000
Product Code: 107144
All lots including and manufactured after 24E0067CB

Oxiris Set Row
Product Code: 973003
All lots including and manufactured after 24F0087
ARTG Number 160473, 354699, 403937, 498166
Unique Device Identifier (UDI) 07332414075613, 07332414075651, 07332414069339, 07332414201241
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action There is a potential for dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit. A dislodged deaeration chamber may lead to an “Air in Blood” alarm on the Prismaflex Control Unit. The alarm may result in a delay/interruption of therapy, and in some cases blood loss due to clotting/failure to return blood manually.

Vantive Australia has received 3 (439 globally) reports related to this problem.
Instructions for Customers/Users - Customers may continue to use the Prismaflex sets by instituting the workaround steps outlined in the Customer Letter.
- Vantive is actively working on a correction for this problem and will advise when available.

Update 05/05/2026: Users are to be aware of the updates to the ARTG Certificates for the oXiris Set. It is now supplied under two ARTG references: 403937 and 498166. Therapy may continue to be performed using the Prismaflex Sets. Vantive is still working on a correction to this problem, and further updates will be provided when it is available.
Action Commencement Date 3/02/2026
Responsible Entity Vantive Pty Ltd
Contact Information 1800 920 133 - Vantive Customer Service