TGA Action ID RC-2026-RN-00024-1
Type of Product Medical Device
Product Name/Description Olympus Cystoscope Outer sheath, 22.5 Fr Sidefire

SKU: WA22810A

Serial Number(s): ALL
ARTG Number 146203
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Olympus has received 8 complaints globally about a Cystoscope Outer sheath with a damaged tip during use of a laser probe. Though the risks associated with the use of a Cystoscope with the laser probe have not changed since the related Product Correction (RC-2024-RN-00883-1) in October 2024, Olympus conducted additional testing, indicating further regulatory actions would be required in order to state compatibility of the cystoscopy sheath with GreenLight laser fibres. Accordingly, Olympus has decided to discontinue the product and recall the device from the market.

Olympus has not received any reported injuries related to this matter.
Instructions for Customers/Users Customers are to:

• Inspect stock immediately and quarantine any stock on hand to prevent further use.
• Complete the Customer Response Form and return it to Olympus.

Once the form has been received, Olympus will reach out to coordinate the return of product and processing of a credit.

This action has been closed out on 30/06/2026
Action Commencement Date 15/01/2026
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations