| TGA Action ID |
RC-2026-RN-00022-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
CleverCut 3V and Flow Cut Sphincterotome
CleverCut 3V Single Use 3-Lumen Sphincterotome V Model(s): KD-V411M, KD-V431M SKU(s): N5411130, N5411230, N5411330, N5411430, N5411530, N5411930, N5412030
CleverCut 3V Single Use Sphincterotome V (Distal Wireguided) with C-Channel Model(s): KD-VC411Q, KD-VC431Q SKU(s): N5777330, N5777530, N5777630, N5777830
KD-401Q-0320 Sphincterotome FlowCut Model: KD-401Q SKU: 5856630
Multiple Lot Numbers |
| ARTG Number |
177705, 326160 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Globally, Olympus has observed an increase in complaints regarding wire deformation in the CleverCut 3V and Flow Cut Sphincterotome devices. Preliminary investigations showed that units manufactured with a thermoforming process, which involves heating the assembly, so the device tip keeps the curved stylet’s shape, are less likely to have an incorrect cutting wire orientation. Devices which did not undergo thermoforming could deform and lose performance. As a result, Olympus is removing lots where the devices may not have been manufactured with the thermoforming process.
The most commonly reported patient harm are delays in initiation or performing procedures. The incorrect orientation of the cutting wire can cause unintended injury to patient anatomy, including bleeding and perforation, requiring surgical or endoscopic intervention to treat. In some cases, improper wire positioning may cause a break and/or detachment, potentially leaving a wire fragment in the patient. This may require additional intervention. |
| Instructions for Customers/Users |
Customers are to: - Immediately inspect and quarantine affected stock to prevent further use.
An Olympus representative will coordinate return of the affected device and process the credit.
This action has been closed out on 07/09/2026 |
| Action Commencement Date |
13/01/2026 |
| Responsible Entity |
|
| Contact Information |
1300 132 992 - Olympus Customer Operations |