TGA Action ID RC-2026-RN-00022-1
Type of Product Medical Device
Product Name/Description CleverCut 3V and Flow Cut Sphincterotome

CleverCut 3V Single Use 3-Lumen Sphincterotome V
Model(s): KD-V411M, KD-V431M
SKU(s): N5411130, N5411230, N5411330, N5411430, N5411530, N5411930, N5412030

CleverCut 3V Single Use Sphincterotome V (Distal Wireguided) with C-Channel
Model(s): KD-VC411Q, KD-VC431Q
SKU(s): N5777330, N5777530, N5777630, N5777830

KD-401Q-0320 Sphincterotome FlowCut
Model: KD-401Q
SKU: 5856630

Multiple Lot Numbers
ARTG Number 177705, 326160
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Globally, Olympus has observed an increase in complaints regarding wire deformation in the CleverCut 3V and Flow Cut Sphincterotome devices. Preliminary investigations showed that units manufactured with a thermoforming process, which involves heating the assembly, so the device tip keeps the curved stylet’s shape, are less likely to have an incorrect cutting wire orientation. Devices which did not undergo thermoforming could deform and lose performance. As a result, Olympus is removing lots where the devices may not have been manufactured with the thermoforming process.

The most commonly reported patient harm are delays in initiation or performing procedures. The incorrect orientation of the cutting wire can cause unintended injury to patient anatomy, including bleeding and perforation, requiring surgical or endoscopic intervention to treat. In some cases, improper wire positioning may cause a break and/or detachment, potentially leaving a wire fragment in the patient. This may require additional intervention.
Instructions for Customers/Users Customers are to:
- Immediately inspect and quarantine affected stock to prevent further use.

An Olympus representative will coordinate return of the affected device and process the credit.

This action has been closed out on 07/09/2026
Action Commencement Date 13/01/2026
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations