| TGA Action ID |
RC-2026-RN-00019-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips MR systems with SW version R11.1 and R12.1
Models: Ingenia 1.5T, Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Ambition S, Ingenia Ambition X, Ingenia Elition S, Ingenia Elition X, MR 7700
Multiple product codes and UDI numbers |
| ARTG Number |
230064, 485786 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Philips has identified potential stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
The MRE viewer PACS indicates around 2 kPa (incorrect stiffness values), which can be associated with healthy tissue, while the general viewing environment in the MR Console viewer indicates almost 8 kPa (correct stiffness values), which can be an indicator of cirrhosis.
No adverse events have been reported. |
| Instructions for Customers/Users |
Customers are to: - Confirm that the system is one of the models listed, and that it has software version R11.1 or R12.1. - Continue to use system(s) in accordance with the IFU. - View stiffness maps in general viewing environment (MR Console viewer) and not MRE viewer PACS, to avoid the potential problem of MRE stiffness value errors.
A Philips representative will contact customers to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the problem. |
| Action Commencement Date |
14/01/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |