TGA Action ID RC-2026-RN-00019-1
Type of Product Medical Device
Product Name/Description Philips MR systems with SW version R11.1 and R12.1

Models: Ingenia 1.5T, Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Ambition S, Ingenia Ambition X, Ingenia Elition S, Ingenia Elition X, MR 7700

Multiple product codes and UDI numbers
ARTG Number 230064, 485786
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified potential stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

The MRE viewer PACS indicates around 2 kPa (incorrect stiffness values), which can be associated with healthy tissue, while the general viewing environment in the MR Console viewer indicates almost 8 kPa (correct stiffness values), which can be an indicator of cirrhosis.

No adverse events have been reported.
Instructions for Customers/Users Customers are to:
- Confirm that the system is one of the models listed, and that it has software version R11.1 or R12.1.
- Continue to use system(s) in accordance with the IFU.
- View stiffness maps in general viewing environment (MR Console viewer) and not MRE viewer PACS, to avoid the potential problem of MRE stiffness value errors.

A Philips representative will contact customers to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the problem.
Action Commencement Date 14/01/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team