TGA Action ID RC-2026-RN-00018-1
Type of Product Medical Device
Product Name/Description Omni Legend PET/CT

All configurations using software version SP5, SP6, SP7, SP8

GTIN: 00195278454218, 00195278592309, 00195278792822, 00195278901439
ARTG Number 400263
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware of a potential intermittent problem on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. In some cases, the artifact can mimic pathology and has the potential to lead to misdiagnosis if undetected. (Note: Standalone CT image acquisition is not impacted by this problem.)

There have been no injuries reported as a result of this problem.
Instructions for Customers/Users A GE representative will contact customers to arrange for the correction.
Action Commencement Date 19/01/2026
Responsible Entity GE Healthcare Australia Pty Ltd