| TGA Action ID |
RC-2026-RN-00017-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Centricity RIS-i configured with Instant Messenger
Software Versions: 7.0.0 – 7.0.6
GTIN: 00840682147170 |
| ARTG Number |
137313 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
GE HealthCare has become aware of a cybersecurity vulnerability affecting Centricity RIS-i configured with Instant Messenger.
When Instant Messenger is installed and running, a malicious actor could exploit this vulnerability to perform unauthenticated Remote Code Execution in Instant Messenger which could potentially impact system availability and/or lead to ability to manipulate data. Instant Messenger is an optional module which is not installed by default.
The vulnerability is not present in RIS-i systems without Instant Messenger, or where Instant Messenger has been disabled. There have been no injuries or unauthorised access to patient data reported to GE HealthCare as a result of this problem. |
| Instructions for Customers/Users |
GE HealthCare will correct all affected products at no cost and a GE HealthCare representative will contact customers to arrange for the correction.
Pending correction by GE HealthCare, customers can continue to use their device by following the instructions below:
Ensure your IT Department disables “Centricity RIS IM Server” service on the server following your Operating System guidelines until product correction is completed.
This action has been closed out on 30/07/2026 |
| Action Commencement Date |
8/01/2026 |
| Responsible Entity |
|
| Contact Information |
1800 659 465 - GE HealthCare Service |