TGA Action ID RC-2026-RN-00011-1
Type of Product Medical Device
Product Name/Description Mondial Monobloc

Serial Numbers: 25C220086, 25C220645 to 25C220654, 25C221448 to 25C221461 inclusive

Product Codes: FWEMB64 (0054258), FWEMB70 (0054261)

Manufacture Dates: Between 27 March 2025 and 16 October 2025
ARTG Number 192110
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Cracks and, in some cases, breakage of the polymer injection-molded leg brace tee component (2390799) have been observed.
Failure of this component may result in unintended movement of the stretcher. Potential unintended lowering or movement of the stretcher, which may cause transport delay or minor injury to patient or operator.
Instructions for Customers/Users Immediately inspect the stretcher for cracks and damage. If damage is observed, remove the stretcher from service until corrected.

If no damage is observed, the stretcher can be used but continual inspection before and after use is required and end users should also ensure a minimum of 2 operators always remain with the stretcher during a patient transport to anticipate and/or control any unintended movement associated with failure of the component.

Trained service technicians will be arranged by Ferno Australia Pty to replace the components.

This action has been closed out on 18/05/2026
Action Commencement Date 7/01/2026
Responsible Entity Ferno Australia Pty Ltd