TGA Action ID RC-2026-RN-00009-1
Type of Product Medical Device
Product Name/Description CERELINK ICP Sensor Basic Kit

Catalogue Number: 826850

Lot Numbers: 7468741, 7482134, 7482137, 7482154, 7482156, 7495593, 7495594, 7495598, 7495664
ARTG Number 363811
Unique Device Identifier (UDI) 10381780520672
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Certain lots of the 14-gauge Tuohy Needle component may potentially have corrosion stains on the surface. The component is included in the CERELINK ICP Sensor Basic Kit.

Potential harms associated with the use of a product with potential corrosion include inflammation, infection, toxic response, and the need for medical intervention beyond the standard of care. If the product has already been used, there is no long-term risk to the patient, and no follow-up is required beyond standard postoperative care.
Instructions for Customers/Users Customers are to:

- Quarantine affected products to prevent further use
- Return the Customer Response Form to Integra as soon as possible

After the form has been received by Integra, customers will be contacted regarding the available replacement product and details for the return of affected goods.

This action has been closed out on 30/07/2026
Action Commencement Date 13/01/2026
Responsible Entity Integra Neurosciences Pty Ltd
Contact Information 1300 550 599 - Integra Customer Service