TGA Action ID RC-2026-RN-00007-1
Type of Product Medical Device
Product Name/Description da Vinci Single Port System Access Port Kits

SP Access Port Small Incision
Part Numbers: 430073
Lot Numbers: K10250522, K10250630, K10250828, K11250821, U10250108, U11241101, U11250219

SP Access Port Large Incision
Part Number: 430075
Lot Numbers: U10250102, U10241107, U10250123, U10250219, K10250529, K10250605
ARTG Number 467226
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action A subset of tray tubs may develop cracks due to a supplier manufacturing problem that caused localised thinning of the plastic during the forming process. Because the tray tub contributes to the sterile barrier, a crack may cause a sterility breach, resulting in the product no longer being sterile at the time of use.

There have been no adverse events reported.
Instructions for Customers/Users Customers are to:
- Review stock and Quarantine affected inventory.
- Complete and return the attached Customer Response Form immediately even if you do not have any affected devices.
- If you identify affected stock, Device Technologies will be in contact with you to arrange their collection.
Action Commencement Date 6/01/2026
Responsible Entity Device Technologies Australia Pty Ltd
Contact Information 1300 338 423 - Device Technologies