TGA Action ID RC-2025-RN-01096-1
Type of Product Medical Device
Product Name/Description ShockPulse-SE Lithotripsy System – Reusable Probes and ShockPulse Lithotripsy Generator

Model Numbers: SPL-SR & SPL-G

Serial Number: CG5047
ARTG Number 343651
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Specific serial numbers of ShockPulse generators may have a mis-wired component. This condition does not impact the intended functionality of the device; however, the miswiring may introduce additional electrical noise on the power supply output. The presence of noise on the ultrasonic input circuit is not compliant with applicable electromagnetic compatibility (EMC) standards and may adversely affect the overall reliability of the system.
In the unlikely event that the ShockPulse Lithotripsy generator’s output becomes reduced, inconsistent, or ceases entirely, there is a potential for procedural delays while troubleshooting or replacing the generator. If a replacement unit is not immediately available, the procedure may need to be cancelled and rescheduled.

OIympus has not received any complaints or reports of serious injury associated with this matter.
Instructions for Customers/Users Following receipt of a completed Response form (supplied to customers by Olympus), Olympus will arrange the return and repair of the affected device at their Repair Centre, after which the device will be returned to the customer.

This action has been closed out on 16/07/2026
Action Commencement Date 2/01/2026
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations