| TGA Action ID |
RC-2025-RN-01089-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Osteobell T OB1110T Biopsy needles
All Osteobell Sizes including: OB0810T, OB0815T, OB0910T, OB0915T, OB1110T, OB1115T, OB1306T, OB1310T
All Batches and expiry |
| ARTG Number |
115608 |
| Unique Device Identifier (UDI) |
NA |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Control Medics have been made aware that overly accentuated manoeuvres or overly forced bending and tilting movements of the needle can cause excessive stress on the cannula. This can cause it to bend or break.
There have been 3 reported cases of needles bending, including 1 instance of breakage of the Osteobell T OB1110T Biopsy needle whilst being removed. |
| Instructions for Customers/Users |
• Customers are advised to read carefully the new IFU before the use of the device to ensure patient safety. • Complete the attached Customer Response form and return to donna@controlmedics.com.au.
This action has been finalised on 13/4/2026. |
| Action Commencement Date |
23/12/2025 |
| Responsible Entity |
|
| Contact Information |
07 3216 2311 - Control Medics |