TGA Action ID RC-2025-RN-01064-1
Type of Product Medical Device
Product Name/Description Fine Round Diamond Bur 1.0MM and 1.5MM

1.0MM Fine Round Diamond Bur
Catalog Number: 5540-011-010
Lot Numbers: 22315017, 22259017, 23053017
GTIN: 04546540700797

1.5MM Fine Round Diamond Bur
Catalog Number: 5540-011-015
Lot Numbers: 23089017, 23062017, 22286017
GTIN: 04546540700803
ARTG Number 232547
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Stryker has reported that there is a potential the device inside the package may not match the product information on the outer label (e.g. label states 5540-011-010 but package contains a 5540-011-015 device or vice versa).

Only the outer packaging is incorrect as the product itself has the correct sizing information lasered onto it.

No injuries related to this problem have been reported to date.
Instructions for Customers/Users Customers are to:
1. Immediately review the Customer Letter and quarantine any affected product they have.
2. Return the Customer Response Form to Stryker with the details of all affected product that is being returned/within their possession.
3. If the customer has loaned, sold or disposed of any of the affected product, they must inform Stryker immediately.

This action has been closed out on 27/08/2026
Action Commencement Date 17/12/2025
Responsible Entity Stryker Australia Pty Ltd