TGA Action ID RC-2025-RN-01058-1
Type of Product Medical Device
Product Name/Description BD Cathena Safety IV Cannula

Product Codes: 386801 386802, 386803, 386804, 386805, 386806, 386807, 386816, 386900

Multiple lot numbers
ARTG Number 243114
Unique Device Identifier (UDI) 00382903868018
Action Type Product Alert
Action Level Hospital
Hazard Classification Class III
Reason for the Action Becton Dickinson (BD) has identified that some Cathena Safety IV Cannulas may exhibit visible excess silicone lubricant at the cannula tip. Silicone lubricant is a standard component of the manufacturing process and is medical-grade, biocompatible, and compliant with ISO 10993-1 requirements. However, in rare cases, excess lubricant may be visible to clinicians upon opening the package or breaking tip adhesion.

This problem may lead to minor delays in therapy initiation if clinicians choose to discard the affected cannula and use an alternative device. Testing confirms that excess silicone remains at the skin level during insertion and does not enter the vascular system.

BD has not received any reports of adverse health events associated with this problem.
Instructions for Customers/Users Customers are to:
- Inspect the cannula tip prior to insertion as per Instructions for Use (IFU).
- If excess lubricant is observed, discard the device and use an alternate cannula.

This action has been closed out on 27/03/2026.
Action Commencement Date 17/12/2025
Responsible Entity Becton Dickinson Pty Ltd
Contact Information 02 4072 0838 - Becton Dickinson