TGA Action ID RC-2025-RN-01054-1
Type of Product Medical Device
Product Name/Description AGILIA SP TIVA

Product code: Z018992

Devices with VSS/VMM dataset lower than 1.2.1
ARTG Number 267873
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action It has been observed that updated Agilia SP TIVA devices are still using non-updated datasets which were created with VSS/VMM versions lower than 1.2.1. This leads to calculation errors and inaccurate dose delivery to patient when using the Alfentanil Scott model.
Instructions for Customers/Users Customers are to:
- Inspect stock immediately and check whether any of the devices have this product code Z018992
- If using datasets of VSS/VMM versions lower than 1.2.1 or are unsure about the dataset version, please cease using the Scott Alfentanil Model.
- Contact the Technical Services team on 1300 695 224 to arrange for the affected devices to be corrected.

This action has been closed out on 14/09/2026
Action Commencement Date 12/12/2025
Responsible Entity Fresenius Kabi Australia Pty Ltd
Contact Information 1300 732 001 - Fresenius Kabi Australia