| TGA Action ID |
RC-2025-RN-01054-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
AGILIA SP TIVA
Product code: Z018992
Devices with VSS/VMM dataset lower than 1.2.1 |
| ARTG Number |
267873 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
It has been observed that updated Agilia SP TIVA devices are still using non-updated datasets which were created with VSS/VMM versions lower than 1.2.1. This leads to calculation errors and inaccurate dose delivery to patient when using the Alfentanil Scott model. |
| Instructions for Customers/Users |
Customers are to: - Inspect stock immediately and check whether any of the devices have this product code Z018992 - If using datasets of VSS/VMM versions lower than 1.2.1 or are unsure about the dataset version, please cease using the Scott Alfentanil Model. - Contact the Technical Services team on 1300 695 224 to arrange for the affected devices to be corrected.
This action has been closed out on 14/09/2026 |
| Action Commencement Date |
12/12/2025 |
| Responsible Entity |
|
| Contact Information |
1300 732 001 - Fresenius Kabi Australia |