| Product Name/Description |
RayStation/RayPlan 10B, 11A, 11B, 12A, 2023B, 2024A, 2024B and v2025 including some service packs
Multiple descriptions, build numbers and UDI numbers. |
| Reason for the Action |
In rare cases, the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs).
This problem occurs for ROI(s) which has no contours and either has a material override defined, or is of type Bolus, Fixation, or Support. It only occurs for ROI(s) that have no contours defined on the image set but have a PrimaryShape. An ROI geometry can reach this state when, e.g., all contours for the ROI have been removed from an image set using a contouring tool, or when a derived ROI is updated or overridden in a way that results in an empty geometry (for example when a derived ROI is the intersection of two ROIs that have no overlap in the image set). After the dose is computed, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
The user will get a warning that dose statistics are not up to date, but when these are recomputed, they are based on the obsolete dose distribution. The treatment plan can be approved and exported with the dose in this inconsistent state. Dose cannot be recomputed unless the user makes another change that invalidates the dose. However, if dose is recomputed or optimisation is redone, the ROI material and geometry will be used correctly. |