TGA Action ID RC-2025-RN-01032-1
Type of Product Medical Device
Product Name/Description basixCOMPAK Inflation Device

Catalog Code: IN4130/K
ARTG Number 322582
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action In rare instances, the handle may detach from the syringe during use due to an insufficient amount of adhesive on the handle, a condition that is not detectable prior to use. Handle detachment during use will result in residual contrast in the balloon, and procedural delay while another syringe is used to aspirate the residual contrast.

There has been an incident where it was reported that the balloon remained inflated after the handle detached, which occurred concurrently with a patient experiencing an adverse event. Simulated use testing has not replicated this failure, specifically the balloon remaining inflated after handle detachment.

The incidence rate for this failure is low at 0.0009%.
Instructions for Customers/Users No product return to Merit is required.

Customers are to follow the steps below to help mitigate the risk of detachment.

a. Holding the cap and barrel together during use may effectively prevent detachment.
b. Visual indicators such as pressure drops on the gauge can alert users to potential failures. Refer to the product Instructions for Use.
c. Simulated use testing showed that handle separation results in flow restoration.
d. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast.

This action has been finalised on 4/11/2026.
Action Commencement Date 11/12/2025
Responsible Entity Merit Medical Australia Pty Ltd