| TGA Action ID |
RC-2025-RN-01022-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
BD Alaris Pump Module model 8100 (pump module) when used with a limited subset of compatible pump infusion sets
Product Codes: 10010454-0006, 11426965-04, 2203-0006, 2420-0007, 2426-0004, 2477-0007, 10942011-04, 2441-0007 |
| ARTG Number |
125916 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
Becton Dickinson (BD) is providing this notice to inform customers of worst-case fluid delivery performance, under certain use cases, for the BD Alaris Pump Module model 8100 (pump module) when used with a limited subset of compatible pump infusion sets.
The subset of infusion sets are those that include in -line filters, back check valves and multiple injections ports. Inclusion of these design features may impact the fluid delivery performance particularly at programmable rates under 1 mL/h and at or under 1 mL of volume.
This data does not impact the performance claims published currently in the BD Alaris System User Manual v12.1.1. |
| Instructions for Customers/Users |
Customers should read the information provided in the Customer Letter reiterating what is in the current IFU (Recommendations and Warnings from the v12.1.1 BD Alaris User Manual) |
| Action Commencement Date |
11/12/2025 |
| Responsible Entity |
|
| Contact Information |
(02) 8875 7000 - BD |