TGA Action ID RC-2025-RN-01022-1
Type of Product Medical Device
Product Name/Description BD Alaris Pump Module model 8100 (pump module) when used with a limited subset of compatible pump infusion sets

Product Codes: 10010454-0006, 11426965-04, 2203-0006, 2420-0007, 2426-0004, 2477-0007, 10942011-04, 2441-0007
ARTG Number 125916
Action Type Product Alert
Action Level Hospital
Hazard Classification Class III
Reason for the Action Becton Dickinson (BD) is providing this notice to inform customers of worst-case fluid delivery performance, under certain use cases, for the BD Alaris Pump Module model 8100 (pump module) when used with a limited subset of compatible pump infusion sets.

The subset of infusion sets are those that include in -line filters, back check valves and multiple injections ports. Inclusion of these design features may impact the fluid delivery performance particularly at programmable rates under 1 mL/h and at or under 1 mL of volume.

This data does not impact the performance claims published currently in the BD Alaris System User Manual v12.1.1.
Instructions for Customers/Users Customers should read the information provided in the Customer Letter reiterating what is in the current IFU (Recommendations and Warnings from the v12.1.1 BD Alaris User Manual)
Action Commencement Date 11/12/2025
Responsible Entity Becton Dickinson Pty Ltd
Contact Information (02) 8875 7000 - BD