TGA Action ID RC-2025-RN-01021-1
Type of Product Medical Device
Product Name/Description Sensis; Sensis Vibe Hemo; Sensis Vibe Combo; AXIOM Sensis Post processing WS; AXIOM Sensis Dedicated SIS Server with HP Flex Pro-C PCs used with the Sensis Vibe application

Product Codes: 10764561, 11007641, 11007642, 06648161, 06648153
ARTG Number 273953
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Sporadic Windows Blue Screen errors have been observed in certain HP Flex Pro-C PCs, causing unexpected restarts and interrupting the Sensis Vibe application. These incidents occur without a clear pattern or trigger.

The root cause is within the Solid-State Drive (SSD) - controller of a certain SSD model which was used within the HP Flex Pro-C PC. These errors delay the start or continuation of examinations.
Instructions for Customers/Users The affected SSD model will be replaced with a different model that has not exhibited the Blue Screen error. Customers will be contacted, and the correction process will commence in February 2026.

In the meantime, please shut down the system and restart it to restore functionality, as described in the Operator Manual.
Action Commencement Date 2/12/2025
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre