| TGA Action ID |
RC-2025-RN-01021-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Sensis; Sensis Vibe Hemo; Sensis Vibe Combo; AXIOM Sensis Post processing WS; AXIOM Sensis Dedicated SIS Server with HP Flex Pro-C PCs used with the Sensis Vibe application
Product Codes: 10764561, 11007641, 11007642, 06648161, 06648153 |
| ARTG Number |
273953 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Sporadic Windows Blue Screen errors have been observed in certain HP Flex Pro-C PCs, causing unexpected restarts and interrupting the Sensis Vibe application. These incidents occur without a clear pattern or trigger.
The root cause is within the Solid-State Drive (SSD) - controller of a certain SSD model which was used within the HP Flex Pro-C PC. These errors delay the start or continuation of examinations. |
| Instructions for Customers/Users |
The affected SSD model will be replaced with a different model that has not exhibited the Blue Screen error. Customers will be contacted, and the correction process will commence in February 2026.
In the meantime, please shut down the system and restart it to restore functionality, as described in the Operator Manual. |
| Action Commencement Date |
2/12/2025 |
| Responsible Entity |
|
| Contact Information |
1800 310 300 - Siemens Customer Care Centre |