TGA Action ID RC-2025-RN-01008-1
Type of Product Medical Device
Product Name/Description Zimmer Air Dermatome and Zimmer Dermatome AN

Item Numbers: 00-8801-001-00 & 88710100

Multiple item lot numbers and serial numbers
ARTG Number 115929
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Worldwide, multiple complaints have been received related to Air Dermatome devices that were cutting beyond the proper depth. Inspections of the returned complaint devices confirmed that affected devices may have a misaligned thickness control bar. If the device has a misaligned thickness control bar from manufacture, it would be expected to manifest such that any potential negative outcomes would be present upon the first few uses.
Instructions for Customers/Users For customers with affected stock, Device Technologies will be in contact to coordinate return and repair in a phased approach. Due to the likelihood that the issue would occur upon the first few uses, affected devices that have not exhibited the issue may continue to be used until requested from Device Technologies to return for repair or preventative maintenance.
Action Commencement Date 8/12/2025
Responsible Entity Zimmer Biomet Pty Ltd