| TGA Action ID |
RC-2025-RN-00994-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Laerdal Compact Suction Unit 4 (LCSU 4 – RTCA)
Part Numbers: 880052,880062,881152
Serial Numbers (SN): 88WWYYXXXX in the range 88132406135 to 88202568266 |
| ARTG Number |
129688 |
| Unique Device Identifier (UDI) |
07045432088519, 07045432088533, 07045432067132 |
| Action Type |
Recall |
| Action Level |
Retail |
| Hazard Classification |
Class III |
| Reason for the Action |
Laerdal has identified that Laerdal Compact Suction Unit 4 (LCSU 4 – RTCA) units manufactured in a limited period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.
The likelihood of electromagnetic interference is low as there is substantial gap between emission limits of the RTCA standard and the susceptibility limits for aircraft. Electromagnetic emissions are significantly below the limit in the standard for electro medical devices (IEC 60601-1-2).
The core functionality of the LCSU 4 (intermittent operation to remove secretions, blood or vomit from a patient’s airway to allow ventilation) is unaffected. |
| Instructions for Customers/Users |
Customers are to: - Check inventory for affected units - Contact Laerdal to arrange a replacement - Retain certain accessories/components to the existing unit (the carry bag, battery, external power supply and wire stand) - Commit to disposal of the affected units
This action has been closed out on 16/04/2026 |
| Action Commencement Date |
27/11/2025 |
| Responsible Entity |
|
| Contact Information |
1800 331 565 - Laerdal Customer Service |