TGA Action ID RC-2025-RN-00992-1
Type of Product Medical Device
Product Name/Description Intracranial Pressure Monitor Pressio 2

Product Code: SOP-PSO-4000 and SOP-PSO-4000R

Multiple Serial numbers
ARTG Number 296621
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Some users have encountered an unintended re-boot of the monitor due to the corruption of the catheter's memory. In this case, a replacement of the catheter is required to continue monitoring the patient.
Instructions for Customers/Users A new software patch has been developed to be used on affected monitors. Users should contact their local sales representative to arrange for the software to be installed.

This action has been closed out on 03/03/2026
Action Commencement Date 27/11/2025
Responsible Entity Life Healthcare Pty Ltd
Contact Information regulatory@ebosmedtech.com - Life Healthcare