TGA Action ID RC-2025-RN-00972-1
Type of Product Medical Device
Product Name/Description Central Monitoring System (BeneVision / HYPERVISOR X)

Workstation PC: HP Pro Mini 400 G9 Desktop PC

Audio Driver: Realtek HD Audio Driver Officially Released by HP

Audio Driver Version: 6.0.9484.1

Multiple serial numbers
ARTG Number 223386
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action When the Workstation of the BeneVision / HYPERVISOR X Central Monitoring system has a specific computer hardware configuration the computer may experience audio playback failure or screen freezing. This is caused by a problem with this version of the audio driver that causes excessive and abnormal system memory consumption.
Instructions for Customers/Users Customers should confirm whether any Central Monitoring System (BeneVision / HYPERVISOR X) are in their facility.

Customers can continue to use the Central Monitoring System. If audio playback failure or screen freezes occurs, they should restart the HP Pro Mini 400 G9 Desktop PC of the workstation.

Mindray service team or authorised service personnel will contact customers as soon as possible for a patch upgrade.

This action has been closed out on 02/04/2026
Action Commencement Date 20/11/2025
Responsible Entity Mindray Medical Australia Pty Ltd
Contact Information 1800 793 011- Paul Woodhouse, Mindray Medical Australia