| TGA Action ID |
RC-2025-RN-00971-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
InteleShare – ProViewer
Release Dates: 3.24.3: 28-Aug-24 to 10-Feb-25, 3.24.4: 20-Nov-24 to 20-Mar-25, 3.25.1: 30 Mar-25 to 8-Sept-25, 3.25.2: 8-Oct-25 to 3-Nov-25 |
| ARTG Number |
461485 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Emergo has identified a defect with the ProViewer contained in InteleShare software versions 3.24.3 - 3.25.2. The defect identified could result in the length measurement tool providing inaccurate values when used in multi-planar reconstruction when rotations are applied. In this scenario, the measurement tool’s text may display inconsistent or incorrect length values.
Inaccurate measurement values could compromise diagnostic accuracy and potentially lead to patient harm, including misdiagnosis or inappropriate follow-up. |
| Instructions for Customers/Users |
Customers are to advise staff to refrain from using measurements in MPR mode after rotating one of the planes.
A software patch will be released and applied to all affected InteleShare servers. There will be no service disruption associated with this release.
This action has been closed out on 07/04/2026 |
| Action Commencement Date |
20/11/2025 |
| Responsible Entity |
|
| Contact Information |
AUNZvigilance@ul.com - Emergo Australia |