TGA Action ID RC-2025-RN-00971-1
Type of Product Medical Device
Product Name/Description InteleShare – ProViewer

Release Dates: 3.24.3: 28-Aug-24 to 10-Feb-25, 3.24.4: 20-Nov-24 to 20-Mar-25, 3.25.1: 30 Mar-25 to 8-Sept-25,
3.25.2: 8-Oct-25 to 3-Nov-25
ARTG Number 461485
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Emergo has identified a defect with the ProViewer contained in InteleShare software versions 3.24.3 - 3.25.2. The defect identified could result in the length measurement tool providing inaccurate values when used in multi-planar reconstruction when rotations are applied. In this scenario, the measurement tool’s text may display inconsistent or incorrect length values.

Inaccurate measurement values could compromise diagnostic accuracy and potentially lead to patient harm, including misdiagnosis or inappropriate follow-up.
Instructions for Customers/Users Customers are to advise staff to refrain from using measurements in MPR mode after rotating one of the planes.

A software patch will be released and applied to all affected InteleShare servers. There will be no service disruption associated with this release.

This action has been closed out on 07/04/2026
Action Commencement Date 20/11/2025
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia