TGA Action ID RC-2025-RN-00966-1
Type of Product Medical Device
Product Name/Description Microstream Advance Neonatal-Infant and Adult-Pediatric Intubated CO2 Filter Line, Microstream Luer Adult-Pediatric Intubated CO2 Filter Line and Microstream Luer Adult-Pediatric Airway Adapter Sampling Line devices

Decription: CO2 MICROSTR AIRADP ADULT/PED, CO2 MICROSTR AIRADP ICU ADT/PE, ETCO2 AIRW ADAPTR MICROSTR N/P, CO2 Microstr.Intubated A/P - E, CO2 Microstr.Intubated A/P - S, CO2 Microstr.Intubated I/N - E

Material Numbers: 7869535, 7869543, 7869550, MP15001, MP15002, MP15003
ARTG Number 123339, 393547
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The manufacturer is updating the Instructions for Use (IFU) for the affected devices.

This IFU update follows receipt of ten (10) reports from customers stating that difficulty or inability in disconnecting the adapter from the patient’s endotracheal tube to perform a procedure was encountered, resulting in a delay to treatment and/or unintended extubation.

Investigation did not identify any anomaly or non-conformance with the product. All devices were manufactured to the specifications set for the product. The existing IFU includes a note instructing users to ensure that the airway adapter can be easily attached and detached from the breathing circuit/tubing during setup before proceeding.
Instructions for Customers/Users Customers are to:
- Be aware of the updated IFU wordings included in the customer letter
- Ensure that the airway adapter can be easily attached and detached from the breathing circuit/tubing during setup as per current IFU
- Continue using the affected devices
- Place the customer letter with the existing IFU as an addendum until the new IFU becomes available.

This action has been closed out on 10/02/2026
Action Commencement Date 21/11/2025
Responsible Entity Draeger Australia Pty Ltd
Contact Information 1800 372 437 - Draeger Customer Service