TGA Action ID RC-2025-RN-00959-1
Type of Product Medical Device
Product Name/Description PALACOS and COPAL Pro Systems

Descriptions: PALACOS R+G pro 40 CE EUR, PALACOS R+G pro 60 CE EUR, PALACOS R+G pro 80 CE EUR, PALACOS MV+G pro 40 CE EUR, PALACOS MV+G pro 60 CE EUR, PALACOS MV+G pro 80 CE EUR, COPAL G+C pro 40 CE EUR, COPAL G+C pro 80 CE EUR

Product codes, 5081273, 5081279, 5081280, 5081281, 5081282, 5081283, 5081284, 5081285
ARTG Number 220925, 309121, 455545
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Heraeus Medical have noted an increase of complaints regarding the breakage of ampoules within their Pro Systems. Difficulties to break both ampoules and breaking failures were reported.

Heraeus Medical found that applying pressure with the palm over the neck of the ampoule casing when flexing it simplifies the ampoule breakage. This advice will be highlighted in the updated Instructions For Use, as well as the product labelling.

No adverse events have been reported to date.
Instructions for Customers/Users - Customers are to read the Customer Letter which details the new Handling Advice.
- A new adaption of the Instruction for Use and labelling of the affected devices including a new pictogram will be introduced.

This action has been finalised on 3/11/2026.
Action Commencement Date 9/12/2025
Responsible Entity Heraeus Medical Australia Pty Ltd
Contact Information 02 8422 6100 - Heraeus Medical