TGA Action ID RC-2025-RN-00958-1
Type of Product Medical Device
Product Name/Description Access 2 Reaction Vessels. An in vitro diagnostic medical device (IVD).

REF: 81901

Lot Numbers: P243943, P243971, P244202 P244344, P244345, P250001, P250002, P250003, P250004, P250005

UDI: 15099590180249
ARTG Number 229315
Unique Device Identifier (UDI) 15099590180249
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Beckman Coulter has identified that affected lots of Access 2 Reaction Vessels (RV) may contain manufacturing deformities. The bottom (round end) of the RV may have a small protrusion or hair like deformity.

The deformed RVs may cause an increased rate of instrument errors, putting the instrument in a Not Ready mode, and delay processing of patient samples.
Instructions for Customers/Users Customers are to:
- Check inventory for affected lots.
- Contact Beckman Coulter for replacement RVs.
- Affected lots may continue to be used until sufficient replacement RVs have been supplied, then the affected lots can be destroyed.

Retrospective review of results is not necessary as this issue does not affect individual patient results.