TGA Action ID RC-2025-RN-00956-1
Type of Product Medical Device
Product Name/Description SynchroMed A820 myPTM application version 1.x and 2.x

GTIN: 00643169771048, 00763000632793
ARTG Number 358090
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action Medtronic has identified a software problem in the SynchroMed A820 myPTM application (app) version 2.x that is installed on the Personal Therapy Manager (PTM).

Delays can occur during actions such as interrogation, resyncing, bolus requests, or unpairing the PTM. The cumulative amount of data that is stored in the PTM after each of these actions causes the myPTM app to slow down. During this time, the progress screen may appear "stuck" at 90% for several seconds or minutes. While the app displays this progress screen, the intended action has already been completed.

This delay may cause confusion for patients, as the app does not readily display the confirmation screen they expect. During the delay the patients may believe the bolus has not been delivered, even though it has already been successfully initiated.

Per design, there is no risk of multiple boluses due to the lock-out feature. In addition, this does not impact delivery of the continuous infusion by the pump.

As of October 2025, Medtronic has received zero (0) reports from patients about delays when using the PTM to interact with their SynchroMed pump in Australia.

Update 05/03/2026: Based on internal testing, Medtronic discovered that this problem is also present in version 1.x. of the application.
Instructions for Customers/Users - Health Care Professionals (HCPs) are to provide a copy of the Patient letter to their patients and maintain a copy of the letter in their records.

- This problem can be temporarily mitigated by clearing data from the system settings. To clear the data, the patient would navigate to Settings - Apps - Patient Data Service - Storage and then select “Clear Data”. However, once cleared, the data will continue to accumulate as the patient interacts with their pump and the app slowness will return, requiring the data to be cleared again. Attachment B of the letter provides additional guidance to clear the data.

Patients are to:
- Identify the software version of the A820 myPTM app as per instructions in Attachment A of the letter.
- Follow the Instructions to temporarily mitigate the delay on the process screen.
- Contact their dedicated healthcare facility to speak with their healthcare team about this problem or if they have any questions.

Medtronic is actively working on a permanent solution and will inform customers as soon as it becomes available.
Action Commencement Date 11/11/2025
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information 0439383614 - James Robinson Medtronic