| Reason for the Action |
Medtronic has identified a software problem in the SynchroMed A820 myPTM application (app) version 2.x that is installed on the Personal Therapy Manager (PTM).
Delays can occur during actions such as interrogation, resyncing, bolus requests, or unpairing the PTM. The cumulative amount of data that is stored in the PTM after each of these actions causes the myPTM app to slow down. During this time, the progress screen may appear "stuck" at 90% for several seconds or minutes. While the app displays this progress screen, the intended action has already been completed.
This delay may cause confusion for patients, as the app does not readily display the confirmation screen they expect. During the delay the patients may believe the bolus has not been delivered, even though it has already been successfully initiated.
Per design, there is no risk of multiple boluses due to the lock-out feature. In addition, this does not impact delivery of the continuous infusion by the pump.
As of October 2025, Medtronic has received zero (0) reports from patients about delays when using the PTM to interact with their SynchroMed pump in Australia.
Update 05/03/2026: Based on internal testing, Medtronic discovered that this problem is also present in version 1.x. of the application. |