| TGA Action ID |
RC-2025-RN-00955-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Rocket Seldinger Chest Drain Kits
Product Codes: R54544-12-SG, R54544-18-SG, R54544-12-PK, R54544-18-PK, R54549-16-PK and R54549-20-PK
UDI: 050552709TF06DXY
Multiple product descriptions and lot numbers |
| ARTG Number |
147548 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Rocket Medical have received two complaints regarding Seldinger Catheters falling out during use.
The Instructions for Use (IFU) (ZDOCK180) for the devices state "The chest tube can now be secured to the skin by using the fixation device R54960-PG-FX or suture or both and is ready for connection to a suitable sealed drainage system such as the Rocket Blue Bottle or Rocket Ambulatory Bag".
The IFU has been updated to include revised guidance regarding the need for a suture.The updated text now states: "The chest tube must now be secured to the skin by suture and is ready for connection to a suitable sealed drainage system such as the Rocket Blue Bottle or Rocket Ambulatory Bag. Further support to the catheter can be provided by the fixation device."
There is also a small label affixed to the outer pack that reads “All You need to use this kit is: 1. A Pair of Gloves 2. Skin Preparation 3. Local Anaesthetic”. This may also potentially mislead users as a suture is not listed.
Future lots will be manufactured without this label. |
| Instructions for Customers/Users |
Customers should review the contents of the letter and updated IFU.
This action has been finalised on 7/4/2026. |
| Action Commencement Date |
12/11/2025 |
| Responsible Entity |
|
| Contact Information |
CustomerServices@rocketmedical.com - Rocket Medical |