TGA Action ID RC-2025-RN-00951-1
Type of Product Medical Device
Product Name/Description ORTHO VISION BioVue Cassettes and Max Analyser for BioVue Cassettes. An in vitro diagnostic medical device (IVD)

Analyser for BioVue Cassettes
Product code: 6904579

Max Analyser for BioVue Cassettes
Product code: 6904578

Sera Anti-N
Product Code: 6904495
ARTG Number 281981, 343753
Unique Device Identifier (UDI) 10758750012831, 10758750012848
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action ORTHO VISION BioVue Analysers may be able to process non-validated tests when they are left on an incorrectly configured regional setting after being reformatted or after new software is loaded. For example, the setting allows the ORTHO Sera Anti-N test to be run on the BioVue Analyzer although it has not been validated for use on the ORTHO VISION BioVue automated platform.

There may be a risk to patient results if ORTHO Sera Anti-N was processed on an analyser where it is not validated, and the ORTHO VISION could potentially produce incorrect results. A false positive result could result in patient injury if an antigen-negative individual is transfused with antigen-positive blood, potentially resulting in haemolytic transfusion reactions.
Instructions for Customers/Users - Do not process the Ortho Sera Anti-N assay on Vision BioVue Analyzers unless it is part of a User Defined Protocol (UDP).
- QuidelOrtho will confirm the correct configuration is present, remotely or via site visit.
Action Commencement Date 11/11/2025
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd
Contact Information 1800 032 359 - Technical Services Centre