TGA Action ID RC-2025-RN-00940-1
Type of Product Medical Device
Product Name/Description OER-AW ENDOSCOPE REPROCESSOR

Material Number: N3044663

Model: OER-AW

All Serial Numbers
ARTG Number 185051
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Olympus received notification of a fire-related incident involving the OER-PRO. After an extensive investigation, the root cause of this incident could not be conclusively determined. Out of an abundance of caution, Olympus is providing Warnings/Cautions reminders associated with all affected Olympus Endoscope Reprocessors.

Harms associated with a fire or thermal event may include user burns or respiratory difficulties. Thermal injuries sustained in facility reprocessing room fires could pose critical or life-threatening risks, particularly in environments where flammable gases, oxygen sources, or combustible materials may be present. Exposure to smoke or fumes emitted may result in respiratory difficulties with prolonged exposure. Additionally, this problem can cause operational disruptions, including delays or cancellations.

There were no patient or staff injuries that occurred during this incident.
Instructions for Customers/Users Customers are to:
- Inspect inventory immediately and identify affected devices.
- Follow the Warnings/Cautions and inspections outlined in the Operation Manual to reduce the potential risk of fire. If any irregularity is observed during the inspections, do not use the equipment and contact Olympus.
Action Commencement Date 4/11/2025
Responsible Entity Olympus Australia Pty Ltd
Contact Information 1300 132 992 - Olympus Customer Operations