| TGA Action ID |
RC-2025-RN-00939-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Meditech FORTIVA
Models: 88IVFT1510LS & 88IVFT1517LS
Lot Number: 20250110
Part Numbers: 88IVFT1510LS & 88IVFT1517LS
Expiry Date: 12/2027 |
| ARTG Number |
447452 |
| Action Type |
Recall |
| Action Level |
Wholesale |
| Hazard Classification |
Class II |
| Reason for the Action |
It has been identified that there may be a problem with the back-check valves on this batch. This means that an infused drug may track into an unused sideline. The potential hazard is that the clinical user may then engage this unused side infusion line, and potentially, unwanted drugs could be infused into the patient. Associated risks would be in line with the unwanted drugs infused, potentially risking/providing a bolus dose.
This market action was undertaken prior to TGA notification. |
| Instructions for Customers/Users |
Customers were instructed to: - Inspect their stock immediately - Return affected stock, or discard it in normal waste.
This action has been closed out on 25/05/2026 |