TGA Action ID RC-2025-RN-00927-1
Type of Product Medical Device
Product Name/Description TPSG2 RoboMarker G2 Tip

Product Code: 61739

Lot Number 1006-C-5
ARTG Number 260688
Unique Device Identifier (UDI) Not yet available
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action IQ Medical has become aware of a lot-specific problem concerning the TPSG2 RoboMarker G2 Tip. When used as per the instructions for use, excess ink may be deposited on the cornea, which may cause blurred lines and temporary corneal staining.

The imprecise line markings may make it more difficult to achieve accurate Toric IOL alignment with the axis of astigmatism of the patient’s cornea, which may result in imprecise or failure of correction of the patient’s astigmatism. Additionally, corneal staining may trigger severe irritation to the cornea, which could potentially lead to corneal inflammation.

This action has been undertaken prior to notification to the TGA.
Instructions for Customers/Users Customers are to:
- Immediately quarantine affected stock on hand to prevent further use
- Return affected stock on hand with the completed response form to IQ Medical.

IQ Medical will provide replacement stock.

This action has been finalised on 10/2/2026.
Action Commencement Date 7/11/2025
Responsible Entity IQ Medical
Contact Information 08 6275 7820 - IQ Medical Customer Service