TGA Action ID RC-2025-RN-00926-1
Type of Product Medical Device
Product Name/Description Xpert CT/NG. An in vitro diagnostic medical device (IVD)

Product Code: GXCT/NGX-CE-10

All batch and lot numbers
ARTG Number 207540
Unique Device Identifier (UDI) N/A
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Cepheid has become aware of a rare genetic mutation event between Neisseria gonorrhoeae and Neisseria meningitidis that caused some Neisseria gonorrhoeae strains to lack target sites for the Cepheid Xpert CT/NG test, leading to false negative results.

A false negative test result for Neisseria gonorrhoeae (NG) may occur. This could lead to transmission of disease and/or disease progression.

This limitation is noted in the product Instructions for Use (301-0234, Sec 18) “Mutations or other changes within the regions of the bacterial genomes covered by the primers and/or probes in the Xpert test may result in failure to detect the target organisms.”
Instructions for Customers/Users Customers are to:
- Monitor for decreases in positivity rate when using the Xpert CT/NG test as the molecular method for detection of N. gonorrhoeae.
- Investigate conflicting laboratory results further and report them to Cepheid.

This action has been closed out on 04/02/2026
Action Commencement Date 31/10/2025
Responsible Entity Cepheid Holdings Pty Ltd
Contact Information 1800 130 821 - Cepheid Holdings