TGA Action ID RC-2025-RN-00925-1
Type of Product Medical Device
Product Name/Description ErgoStars with smoothbore hoses incl. Sets and Packs using those products

ErgoStar CM 40
Part No. MP01840

ErgoStar CM 45
Part No. MP01845

ErgoStar CM 55
Part No. MP01855

ErgoStar CM 60
Part No. MP01860

CombiStar Plus Filter + HME straight
Part No. MP12060

Pack2Go Evita invasive Vpack CO2 Cuvette
Part No. MP20370

Pack2Go Evita invasive Vpack
Part No. MP20371
ARTG Number 287190, 374261, 461205
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Dräger has become aware of cases in which cracks have formed in the hose of some specific ErgoStar catheter mounts. No patient harm was reported in these events.

In case such cracks form during use, leakage will occur. Depending on the size of the leakage and the selected ventilation pressure, the ventilator / anaesthesia device can compensate for this. If the leakage cannot be compensated for, ventilation is restricted, and the patient may experience desaturation/hypoxia. Ventilators according to ISO 80601-2-12 and anaesthesia devices according to ISO 80601-2-13 will generate visual and audible alarms according to the set alarm limits.
Instructions for Customers/Users Customers should identify affected products in their inventory and cease use of the ErgoStar.

All components in the affected Pack2Go except for the ErgoStar may be used properly.

Dräger recommends the alternative product MP01850 ErgoStar CM 50 as a suitable replacement.
This product is not affected by the current issue and can replace stock free of charge through the Dräger representative. Please note that it is not the identical item, but a comparable alternative from Dräger's portfolio.

This action has been finalised on 17/4/2026.
Action Commencement Date 30/10/2025
Responsible Entity Draeger Australia Pty Ltd
Contact Information 1800 372 437 - Customer service