TGA Action ID RC-2025-RN-00917-1
Type of Product Medical Device
Product Name/Description Full Body UVB Therapy Unit

Software-Version: xx_2.0.39/2.01

Part No. 4557 N-LINEpro UVB 311nm
Part No. 4558 N-LINEpro UVA
Part No. 4559 N-LINEpro Universale

Serial Numbers: 455723-0000107, 455725-0000202, 455725-0000101
ARTG Number 371303
Unique Device Identifier (UDI) 4065258010054
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action MEDlight GmbH reports a rare case of rapid, unusual user actions including the entry of an mJ value instead of a J value, the non-observance of visual and acoustic warnings regarding both the dose and the duration of treatment (manual override of a soft stop by the user) and the non-monitoring of the application by the user.

This operating sequence could have led to an unintended UV overexposure of the patient, which could result in patient harm including burns.
Instructions for Customers/Users Customers are to:
- Enter dose in the software in J/cm² and refrain from selecting doses in the unit range mJ/cm².
- Observe and check warning notices.

- MEDlight will provide a software update for the affected products, which will prevent treatment from starting even if the warning messages are ignored in situations similar to the one described above (hard stop).

This action has been closed out on 23/06/2026
Action Commencement Date 29/10/2025
Responsible Entity Allied Biotech Australia Pty Ltd