TGA Action ID RC-2025-RN-00911-1
Type of Product Medical Device
Product Name/Description Nutricia Flocare Infinity III enteral feeding pump AC plug Head

Batch Numbers: KRDD32M, KRDD387, KRDD412, KRDD3BL, LYID43D, LYID42N, LYID43D, LYID459, LYID46T, LYID48D

Multiple batch, serial and UDI numbers
ARTG Number 119667
Unique Device Identifier (UDI) Plug head - N/A
Action Type Product Correction
Action Level Consumer
Hazard Classification Class II
Reason for the Action Nutricia has received an internal report indicating that the AC plug heads on up to 10 batches of Flocare Infinity III pump chargers may be susceptible to breakage or disconnection when subjected to axial force, resulting in potentially unsafe conditions and faulty plug performance.

No adverse events have been reported to date in Australia.
Instructions for Customers/Users Customers are to
- Immediately inspect the AC plug heads of affected batch numbers
- Isolate any AC plug heads that are loose, broken, or damaged to prevent further use and request a replacement.
- Continue using pumps with an alternative Infinity charger or resume use once the damaged plug head has been replaced.
- Continue to follow Instructions for Proper Handling of the Flocare Infinity III Plug Head.
Action Commencement Date 4/11/2025
Responsible Entity Nutricia Australia Pty Ltd
Contact Information 1800 060 051 - Nutricia Australia