TGA Action ID RC-2025-RN-00909-1
Type of Product Medical Device
Product Name/Description ELITA Femtosecond Laser System

Product Code: 0155-1001

Serial Number: 0623-11061
ARTG Number 409564
Action Type Product Alert
Action Level Hospital
Hazard Classification Class III
Reason for the Action AMO Australia identified that at high altitudes above 1.1 km, the vacuum pump may not meet the established pressure specifications. Therefore, the system will not meet the specification for operation at the maximum altitude of 2.4km, as stated in the product’s installation instructions.

There is no expected performance impact to the vacuum pump at or below 1.1 km altitude; however, if the device was installed at an altitude greater than 1.1 km, the system vacuum pump may not be able to reliably generate adequate vacuum for the system to run, and the system will not function.

There is no impact to patient safety, as the ELITA System’s software design prevents use if vacuum checks fail.
Instructions for Customers/Users Customers are to:
- Review the Operator Manual and customer letter (sent to affected customers) to ensure understanding.

This action has been finalised on 20/8/2026.
Action Commencement Date 24/10/2025
Responsible Entity AMO Australia Pty Ltd
Contact Information 1800 266 111 - AMO Australia