| Reason for the Action |
AMO Australia identified that at high altitudes above 1.1 km, the vacuum pump may not meet the established pressure specifications. Therefore, the system will not meet the specification for operation at the maximum altitude of 2.4km, as stated in the product’s installation instructions.
There is no expected performance impact to the vacuum pump at or below 1.1 km altitude; however, if the device was installed at an altitude greater than 1.1 km, the system vacuum pump may not be able to reliably generate adequate vacuum for the system to run, and the system will not function.
There is no impact to patient safety, as the ELITA System’s software design prevents use if vacuum checks fail. |