| TGA Action ID |
RC-2025-RN-00895-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Bicarbonate Reagent. An in vitro diagnostic medical device (IVD).
REF: OSR6137, OSR6237, OSR6637, OSR6190 4 x 25mL R1, 4 x 740 Tests on AU5800, OSR62904 x 50mL R1, 4 x 1440 Tests on AU5800
Multiple lot numbers and expiry dates. |
| ARTG Number |
185654, 200019 |
| Unique Device Identifier (UDI) |
UDI number is 15099590060206 for Part Number : OSR6637, UDI number is 15099590020187 for Part Number : OSR6237, UDI number is 15099590010348 for Part Number : OSR6137, UDI number is 15099590020330 for Part Number : OSR6290, UDI number is 15099590010867 for Part Number : OSR6190 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Beckman Coulter has received complaints for elevated results generated using Bicarbonate reagent when there are high levels of Lactate Dehydrogenase (LDH) in the test sample. The affected lots of Bicarbonate reagent may generate Bicarbonate results that are biased high due to interference with LDH in the test sample. The level of LDH interference is not consistent across lots of Bicarbonate reagents.
This problem does not affect Bicarbonate samples with LDH levels within the normal reference range for an adult patient. |
| Instructions for Customers/Users |
Customers are to discontinue use and dispose all Bicarbonate OSR6x37/OSR6x90 lots listed in Table A in the PROBLEM section of the customer letter, following laboratory protocols and local regulations.
For Bicarbonate OSR6x37/OSR6x90 lots listed in Table B in the PROBLEM section of the customer letter - if elevated LDH levels are suspected, please assess results in conjunction with patient’s LDH values.
Contact your local Beckman Coulter representative for replacement or re-imbursement. |
| Action Commencement Date |
20/10/2025 |
| Responsible Entity |
|
| Contact Information |
1800 060 878 - Beckman Coulter |