| TGA Action ID |
RC-2025-RN-00882-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
3gAllergy Specific IgE Universal Kit. An in vitro diagnostic medical device (IVD).
Lot Numbers: 169, 170, 171, 172
Expiry Date: 22/07/2026
Catalogue Number: L2KUN6
UDI: 00630414962269
Material Number: 10380875 |
| ARTG Number |
179719 |
| Unique Device Identifier (UDI) |
00630414962269 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Siemens Healthineers has identified that barcode orientation on the 3gAllergy Specific IgE Universal Kit causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
There is the potential for patient sample results to be inaccurate for any allergen loaded in the same allergen holder wedge. Worst case, a false non-reactive result when truly reactive may lead to lack of avoidance of the allergen and possible anaphylaxis. |
| Instructions for Customers/Users |
Customers are to: - Review the customer letter with a Medical Director to determine the appropriate course of action, including for any previously generated results. - Discontinue use of and discard the affected kit lots. - Contact Siemens Healthineers to arrange replacement stock. |
| Action Commencement Date |
17/10/2025 |
| Responsible Entity |
|
| Contact Information |
1800 310 300 - Siemens Customer Care Centre |