TGA Action ID RC-2025-RN-00866-1
Type of Product Medical Device
Product Name/Description MAS Omni CARDIO Liquid Assayed Integrated Cardiac Control. An in vitro diagnostic medical device (IVD)

Expiry date: 30/04/2027

Catalog Numbers: CDFOCRD-UL, CDFOCRD-L, CDFOCRD-101, CDFOCRD-202, CDFOCRD-303

Lot Numbers: OCRD2704U, OCRD2704L, OCRD2704, OCRD27042, OCRD27043

UDI: 00884883011946, 00884883011953, 00884883011960, 00884883011977, 00884883011984
ARTG Number 192625
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action MAS Omni CARDIO Liquid Assayed Integrated Cardiac Control has been reported to have vial-to-vial variability when using the control with high sensitivity Troponin I assays.

No other analytes are reported to be showing variability between vials.

If there is a delay in reporting of patient results, there is a remote probability of a serious adverse event in the overall population using the device. There is no risk of incorrect results being generated.
Instructions for Customers/Users Labs may continue to use the control if it meets their performance requirements.

If the variability of the material exceeds user's laboratory’s specification, dispose of the material. Contact the Thermo Fisher Scientific Account Manager or the Customer Care Team at 1300 735 292 or email ANZcustomerservice@thermofisher.com

This action has been finalised on 7/4/2026.
Action Commencement Date 13/10/2025
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd
Contact Information 1300 735 292 - Customer Care Team