TGA Action ID RC-2025-RN-00864-1
Type of Product Medical Device
Product Name/Description CARESCAPE B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait Vital Signs Monitor

Multiple product names, REF numbers and GTIN's
ARTG Number 115304, 118982, 378143, 496489
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait Vital Signs Monitors when powered by certain batteries. Affected batteries can lose their capability to hold charge and provide power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognising patient status changes requiring treatment.

There have been no injuries reported to GE HealthCare as a result of this problem.
Instructions for Customers/Users If possible, use a monitor with a battery that is not within the affected battery population listed. However, pending correction by GE HealthCare, customers can continue to use a monitor with affected batteries by following the instructions in the User Manual.

GE HealthCare will correct all affected products at no cost. A GE HealthCare representative will contact affected customers to arrange for the correction.

This action has been finalised on 03/11/2026.
Action Commencement Date 15/10/2025
Responsible Entity GE Healthcare Australia Pty Ltd