| TGA Action ID |
RC-2025-RN-00864-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
CARESCAPE B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait Vital Signs Monitor
Multiple product names, REF numbers and GTIN's |
| ARTG Number |
115304, 118982, 378143, 496489 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE B450, CARESCAPE ONE, B1x5M/B1x5P VSP3.0, B1x5M/B1x5P VSP4.0 Patient Monitors and Portrait Vital Signs Monitors when powered by certain batteries. Affected batteries can lose their capability to hold charge and provide power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognising patient status changes requiring treatment.
There have been no injuries reported to GE HealthCare as a result of this problem. |
| Instructions for Customers/Users |
If possible, use a monitor with a battery that is not within the affected battery population listed. However, pending correction by GE HealthCare, customers can continue to use a monitor with affected batteries by following the instructions in the User Manual.
GE HealthCare will correct all affected products at no cost. A GE HealthCare representative will contact affected customers to arrange for the correction.
This action has been finalised on 03/11/2026. |
| Action Commencement Date |
15/10/2025 |
| Responsible Entity |
|
| Contact Information |
1800 659 465 - GE HealthCare |