| TGA Action ID |
RC-2025-RN-00852-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Perfect Control Cells 0.8% 2x10mL and Perfect Control Cells 3% 2x10mL. An in vitro diagnostic medical device (IVD)
Perfect Control Cells 0.8% 2x10mL Product code: 729901
Perfect Control Cells 3% 2x10mL Product code: 727806
Batch number: 25807.01
Expiry: 03/11/2025 |
| ARTG Number |
235293 |
| Action Type |
Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
During pre-acceptance of the product, a customer reported a false positive antibody (IAT) result obtained for Cell 2 (rr) when using a Weak Anti-D Control. Further tests identified a DAT positive result for Cell 2. Due to this status false positive results may occur when the product is used in the Indirect Antiglobulin Test (IAT). Cell 1 is not affected.
There is a low risk that an unexpected positive IAT result for Cell 2 of the Perfect Control Cell product will need additional investigation prior to verifying the status of the test system.
The risk does not apply to the use of the batch for Blood Grouping tests. |
| Instructions for Customers/Users |
- Customers are to be reminded of the Test Method section of the IFU for Perfect Control Cells:
As test methodologies will vary significantly between laboratories, the cells should be verified for use on the validated test methods employed by the individual laboratory by its validation procedures.
- For an alternative for the IAT, an equivalent blood group O positive RhD Negative (rr) Cell can be used as a substitute for Cell 2 (e.g. from Perfect Panel 11 Cell 0.8% 11x10ml or Perfect Panel 16 Cell 0.8% 16x4ml).
- Unaffected product will be manufactured and supplied as soon as it is available.
This action has been closed out on 04/02/2026 |