TGA Action ID RC-2025-RN-00846-1
Type of Product Medical Device
Product Name/Description A9 Anaesthesia System and V90 electronic vaporiser used with A9 Anaesthesia system

Multiple serial numbers
ARTG Number 158607
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Mindray has identified the following potential problems with the A9 anaesthesia system and the V90 electronic vaporiser used with the A9 anaesthesia system.

1. V90 electronic vaporiser leakage: During the process of filling the anaesthetic agent into the vaporiser frequently applying excessive downward pressure on the anaesthetic agent bottle may damage the internal structure of the filling port. This damage may lead to improper sealing of the filling port after the removal of the bottle and may result in anaesthetic leakage.

2. A9 Anaesthesia System Occasional alarm of “vaporiser output abnormal” is triggered during operation, but the actual concentration is normal. Due to the occasional abnormal zeroing of the internal AG module of the device, the monitoring value is on the low side and triggers the false alarm; the monitoring value of the internal AG module is only used to judge whether the output concentration of the vaporiser is normal or not, the monitoring value will not be displayed on the screen, so it will not mislead the customers and will not affect the actual concentration of anaesthetics output from the vaporiser.

3. One-Step Lung recruitment of the A9 anaesthesia system abnormality: After using the One-Step Lung recruitment function, it may fail to be activated again under a specific operation path. Ventilation and other functions are normal at this time, and the user can manually set the ventilation parameters to achieve lung recruitment.
Instructions for Customers/Users Problem 1:
- The V90 electronic vaporiser can still be used as intended, however, please note, when adding aesthetic agent, to press the bottle slowly downward and avoid using excessive force.
If there is smell of anaesthetic when using it, either replace it with another unaffected V90 vaporiser or use intravenous anaesthesia.

- The affected V90 vaporisers will be replaced with corrected ones equipped with an improved filling port structure.

Problem 2:
- The A9 anaesthesia system can still be used as intended. When the alarm “Vaporiser Output Abnormal” occurs, if the monitoring value of anaesthetics at the patient side is normal, it is a false alarm triggered by the deviation of the internal AG module in calibrating the zero, and the actual output concentration of the vaporiser is normal. The false alarm will disappear by itself after the internal AG module calibrates the zero again successfully, or after manually restarting the A9 anaesthesia system.

- A software update will be implemented.

Problem 3:
- When the One-Step Lung recruitment function of the A9 system cannot be activated, the user can manually set the ventilation parameters to achieve the lung recruitment. When powering on the A9 anaesthesia system again, the One-Step Lung recruitment function will return to normal.

- A software update will be implemented.

This action has been closed out on 09/06/2026
Action Commencement Date 21/10/2025
Responsible Entity Mindray Medical Australia Pty Ltd
Contact Information 1800 793 011- Mindray Medical Australia